HTHP Stirred/Static Corrosion Instrument is used for corrosion test of small metal sample.

Specifications:

*reaction vessel fluid capacity 1000ml

*rotary sample rack: max.4 corrosion samples or one rock core sample at the same time

*pipe line size: 6.35mm OD

*max. test pressure and temperature: 200MPa at 200oC

*sulfuric acid concentration: 28%

*pressure transducer precision: -/+0.25% of full scale(+/-0.9MPa)

Safety device:

*Temperature limit

The limit switch will turn off the heater supply when temperature is over the set temperature.

*Pressure limit device

Rupture disk is connect to the bottom of raction vessel, pressure is about more than 20MPa, the liquid discharges to downward safely.